2000-11-15
Got SAF form and enabled beamline.
Johnny is using a new pumping stand for the mono chamber (pair of Si crystals).
The old pump caused to much vibrations. As the pumping stand is located so that
it can be kicked out by X23A users, checking the mono tank vacuum would be
pretty wise.
There are a few important changes in the handling of the lockin:
On starting XDAC turn on Digital Feedback in the Mono setup menu.
Changing the gain is not necessary any more. (That is what the feedback does.)
Adjust mirrors in the middle of the scan, then return to the beginning and set
BIAS to zero there.
We enabled the beamline and let it warm up.
A beamdump occurs just before I wanted to start the first edge scan on a V
foil. They had to access the ring tunnel and repair a cooling water supply.
After the injection (15:30) I start the edge scan:
| Filename | sample | comments | ranges | stepsize | # scans |
| 15Nov00.001 | V foil | edge scan | -100, 100 | 2 | 1 |
| calibrated mono to V edge | |||||
| 15Nov00.002 | V foil | near edge scan | -30, 30 | 1 | 1 |
| calibrated mono to V edge |
Another beamdump at 16:30!
I checked which motor numbers are used:
The correct names are Sample-Horiz, Sample-Vert and Sample-Rot with the numbers
3,4,1
mounted sample 1156 K2 S... with E vector parallel c axis
| Filename | sample | comments | ranges | stepsize | # scans |
| 15Nov00.003 | V2O3 1156 K2 |
motor scan Z (wrong slice number in data comments) |
-4, 4 | .1 | 1 |
| calibrated motor | |||||
| 15Nov00.004 | motor scan Y | -2, 2 | .1 | 1 | |
| calibrated motor | |||||
| 15Nov00.005 | motor scan Z | -2, 2 | .1 | 1 | |
| recalibrated motor | |||||
| 15Nov00.006 | edge scan Gain wrong on Io, has to be 1e7 |
-100, 100 | 1 | 1 |
After the third beamdump for this afternoon, I went to pick up Mihail and
Roman from the station.
The beam returned at 8.30 p.m. We let the mirrors warm up again and started
with a readjustment of the sample:
| Filename | sample | comments | ranges | stepsize | # scans |
| 15Nov00.007 | V2O3 1156 K2 |
motor scan Z | -2, 2 | .1 | 1 |
| calibrated motor | |||||
| 15Nov00.008 | motor scan Y | -2, 2 | .1 | 1 | |
| calibrated motor | |||||
| 15Nov00.009 | edge scan | -100, 100 | 1 | 1 | |
| recalibrated mono | |||||
| 15Nov00.010 | EXAFS scan Gain wrong on Io, has to be 1e7 |
-200, -10, 30, 16k | 2,2,2 | 2, .5, .05k | |
| mono recalibrated (from 4551eV) | |||||
| 15Nov00.011 15Nov00.012 |
Gain wrong on Io, has to be 1e7 |
2000-11-16
Johnny obvoiusly was here lat at night (11 p.m.?). He left me a note, telling
that he changed the scale on the oszi and turned on Io plotting in the XDAC.
At the beginning of the third scan it lost lock and Johnny retuned the optics
and started the scans over.
He reduced sensitivity from 500mV to 1V on the lockin amp. as is was in
overload.
He also mentioned that Io looked very big to him and suggested that I should
add som He to Io or to shut down the slits a bit . Otherwise Io might not be
linear.
Gain on Io is quite high (1e9), so probably thats not necessary. I have to talk
to Johnny about that today. Closing the slits should not hurt in any case with
my sample size.
| Filename | sample | comments | ranges | stepsize |
| 15Nov00.013 - 017 | V2O3 1156 K2 |
EXAFS scans #017 seems to be empty Gain wrong on Io, has to be 1e7 |
-200, -10, 30, 16k | 2, .5, .05k |
Emergency ring access seems to take longer than they thought. They're starting the injection at 8:45.
At some point I changed the Gain reading for Io in the XDAC to its correct
value of 1e7
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.001 | V2O3 1156 K2 |
time scan 2 check lin. of Io: 1st part: no Al foil 2nd part: 1 Al foil 3rd part: 2 Al foils |
0, 1000 | 1 |
Noticed a hole in the aluminized mylar foil at the end of the beamline
extension in the hutch. fixed it.
Reduced exit slits to 1 x 1 mm.
I_t is significantly reduced. Doing motor scans again:
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.002 | V2O3 1156 K2 |
motor scan Z no change in Z position wrong user comments in data file |
-2, 2 | .1 |
| 16Nov00.003 | motor scan Y only small change wrong user comments in data file |
|||
| 004 | time scan again | 0, 1000 | 1 |
No good result: linearity changes 5-10%, Johnny says tey got it to as good
as .5%.
So I open up the slits again, increasing the I_t.
Using mixture of N2 and He for Io chamber now. Flowmeter shows 40 for N2 and 10
for He. Io reduced from 3.05 to 2.65. Testing linearity again:
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.005 | V2O3 1156 K2 |
time scan | 0, 1000 | 1 |
A propos gas flows:
| Device | He | N2 | Ar | |||
| Beamline window | red | 17 | - | - | - | - |
| Beamline extension | green | 90 | - | - | - | - |
| Io (still testing) | blue | 30 | white | 10 | - | - |
| I_t | - | - | gray | 40 | - | - |
| I_Lytle (not in use) | - | - | - | - | yellow | (33) 0 |
| (not in use) | brown | |||||
| (not in use) | black | |||||
Yesterday we connected the wrong hoses and were blowing N2 into the hutch
instead of into the ion chambers. That problem is fixed, now.
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.006 | V2O3 1156 K2 |
time scan lost beam in the region of 200 sec 0,1,2 layers of Al foil |
0, 1000 | 1 |
| 007 | check for stabilizing Io changed He/N2 to 20/20 at 144 |
|||
| 008 | continuing from scan 007, without glitch | |||
| 009 | another time scan changed He/N2 to 25/15 at 80 |
|||
| 010 | dito, 009 was slowly stabilizing at 1000 |
I decide to leave the gas flows constant now. They are:
| Device | He | N2 | Ar | |||
| Beamline window | red | 17 | - | - | - | - |
| Beamline extension | green | 90 | - | - | - | - |
| Io (still testing) | blue | 25 | white | 15 | - | - |
| I_t | - | - | gray | 40 | - | - |
| I_Lytle (not in use) | - | - | - | - | yellow | (33) 0 |
| (not in use) | brown | |||||
| (not in use) | black | |||||
Doing another linearity check. Hopefully it's the last one:
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.011 | V2O3 1156 K2 |
time scan 4 linearity check | 0, 1000 | 1 |
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.012 | V2O3 1156 K2 |
time scan 4 linearity check 2 foils, 1 foil, no foil |
0, 1000 | 1 |
Final gas flow table:
| Device | He | N2 | Ar | |||
| Beamline window | red | 17 | - | - | - | - |
| Beamline extension | green | 90 | - | - | - | - |
| (not in use) | blue | 0 | ||||
| Io | - | - | white | 40 | - | - |
| I_t | - | - | gray | 40 | - | - |
| I_Lytle (not in use) | - | - | - | - | yellow | (33) 0 |
| (not in use) | brown | |||||
| (not in use) | black | |||||
During the overnight scans, the energy has shifted by MORE THAN 100 eV. I
corrected that this morning, but rather do another edge scan:
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.013 | V2O3 1156 K2 |
edge scan | -100, 100 | 1 |
| recalibrated mono |
Edge was off 50 eV
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.014 | V2O3 1156 K2 |
motor scan Z | -2, 2 | .1 |
| 015 | motor scan Y | |||
| 016 | EXAFS scan aborted after edge: was shifted over 10 eV |
-200, -10, 30, 16k | 2, .5, .05k | |
| 017 | EXAFS scan | -200, -30, 30, 16k | 2, .5, .05k |
Switched sample to have E perpendicular c now. Doing motor scans
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.018 | V2O3 1156 K2 |
E perp c now motor scan Z |
-4, 4 | .1 |
| 019 | motor scan Y | |||
| did a burnpaper check and moved stage horizontally | ||||
| 020 | motor scan Z | |||
| 021 | motor scan Y | |||
| 022 | EXAFS scan E perp c killed by PP after edge |
-200, -30, 30, 16k | 2, .5, .05k | |
| mono recalibrated | ||||
| 023 | EXAFS scan E perp c |
|||
| MONO RECALIBRATED | ||||
| 024 | dito | |||
| MONO RECALIBRATED | ||||
| 025 | dito |
Mounted new sample 608 E8 Rest w/ E perpendicular c (??? Sample orientation has to be rechecked, Perpendicular orientation results from comparison with spectra of other samples). Using graphite for thermal contact and wiring for resistance measurement.
Starting with motor scans:
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.026 | V2O3 608 E8 R |
motor scan Z | -2, 2 | .1 |
| 027 | motor scan Y | |||
| 028 | edge scan | -100, 100 | 1 | |
| RECALIBRATED MONO | ||||
| 029 | EXAFS scan Obviously E perp. c |
-200, -30, 30, 16k | 2, .5, .05k | |
| 030 | EXAFS scan Obviously E perp. c |
After dinner we decide to go on with Mike's and Roman's experiment.
We tilt the Displex out of the way and put up their Ti foil.
| Filename | sample | comments | ranges | stepsize |
| 16Nov00.031 | Ti foil | edge scan no current |
-100, 100 | 2 |
| recalibrated mono | ||||
| 16Nov00.032 | near edge scan no battery |
-30, 30 | 0.3 | |
| 16Nov00.033 | near edge scan 1 battery connected |
|||
| 034 | without battery again | |||
| 035 | with battery | |||
| 036 | with 4 batteries |
2000-11-17
Just looked up the transition temperature of charge 608: 141 K on cooling, 148
K on warming up.
We have set up the Displex. I took my chance and spilled some cooling water on
Roman :-(
The forepump runs quite smoothly and does not seem likely to fall off the
stool.
Resistance of sample is 76 Ohms @ 220 K
Looking for sample with motor scans:
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.001 | V2O3 608 E8 Rest |
motor scan Z | -2, 2 | .1 |
| 002, 003, 004 | motor scans Y | different ranges | different stepsizes | |
| 005 | motor scan Z | -4, 4 | .2 | |
| 006 | motor scan Y | -2, -2 | .1 | |
| 007 | motor scan Z | |||
| 008 | edge scan | -100, -30, 30, 100 | 2,1,2 |
Found the sample, calibrated the edge, retuned the optics.
Starting an EXAFS scan now
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.009 | V2O3 608 E8 Rest |
EXAFS scan E perp c (?) 200 K, 81.4 Ohm |
-200, -30, 30, 16k | 2, .5, .05k |
| mono recalibrated (5452 -> 5465) | ||||
| 010 | 200 K, 55 Ohm (new battery in Multimeter) |
Johnny tried something on the mono motors which did not work yet. He
restored the old settings and I take another scan after retuning the optics.
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.011 | V2O3 608 E8 Rest |
EXAFS scan E perp c (?) 200 K, 55.9 Ohm ABORTED |
-200, -30, 30, 16k | 2, .5, .05k |
In each of the scans there are strange structures in the preedge region (
around 5300 eV) that change slightly from scan to scan. But that was not the
reason why I aborted the scan. I noticed that the sample position has changed
and the I_t signal was very low.
So doing motor scans again:
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.012 | V2O3 608 E8 Rest |
motor scan Z | -2, 2 | .1 |
| 013 | motor scan Y | |||
| 014 | edge scan Changed gain on I_t to 1e10, but forgot to measure offsets |
-50, 50 | 1 | |
| calibrated mono (5432 -> 5465) |
Measured offsets now, after changing gain on I_t to 1e10. Starting an EXAFS
scan:
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.015 | V2O3 608 E8 Rest |
EXAFS scan E perp c (?) 200 K, 56.1 Ohm I_t detector went to nearly 11 V at the very end of the scan. |
-200, -30, 30, 16k | 2, .5, .05k |
| 016 | testing alternative mono drive (energy driving wrong way) | |||
| 017 | dito | |||
| 018 | edge scan 170 K, 64.1 Ohm |
-100, 100 | 2 |
Reset I_t gain to 1e9 and measured offsets (really did it this time befor
the first scan!).
Probably low gain is better for the scan, even if signal is low. But Johnny
says that linearity is not guaranteed abov 8 Volts - and I was reaching 11 V!
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.019 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 155K, 70 Ohm |
-200, -30, 30, 16k | 2, .5, .05k |
| calibrated mono (5453 -> 5465) | ||||
| 020 | motor scan Y | |||
| 021 | motor scan Z | |||
| 022 | EXAFS scan E perp. c (?) 155K, 70.5 Ohm |
|||
| mono recalibrated ( 5447 -> 5465 ) | ||||
| 023 | EXAFS scan E perp. c (?) 155K, 70.1 Ohm |
|||
| 024 | dito | |||
| mono recalibrated (5429 -> 5465) |
Went to 149 K (73 Ohm), maybe the transition occurs at a temp. a bit lower
than measured in Augsburg.
Beam dumped (2:45). I "stand by".
New injection, new photons, new alignment:
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.025 | V2O3 608 E8 Rest |
motor scan Z | -2,2 | .1 |
| 026 | motor scan Y |
Just talked to Joe. He will have some beamtime in Feb. or March.
He looked at my k-space data of the 155 K spectra and was suspicious about the
sharp peak at 12 A^-1. I will rotate the sample a few degrees after the next
scan and see if that changes (and is a Bragg peak then) or if it's really in
the data.
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.027 | V2O3 608 E8 Rest |
EXAFS scan E perp c (?) 149 K, 73.5 Ohm |
-200, -30, 30, 16k | 2, .5, .05k |
| recalibrated mono from 5440 | ||||
| rotated sample to 4 units of "sample rot" motor (should be 4 or 8 deg.), was at 0 units before moving. | ||||
| 028 | motor scan Y | |||
| moved to -.3 but did not recalibrate motor | ||||
| 029 | motor scan Z | |||
| moved to +.3 but did not recalibrate motor | ||||
| 030 | EXAFS scan 149 K, 73.7 Ohm |
|||
| recalibrated mono from 5447 | ||||
| rotated sample to -4 units | ||||
| 031 | motor scan Z | |||
| moved to .4 w/out calibration | ||||
| 032 | motor scan Y | |||
| moved to -.2 w/out calibration | ||||
| 033 | EXAFS scan dito |
|||
| recalibrated mono from 5450 |
I believe that the sharp structure is just noisy. All three scans look quite
the same.
Returning to 0 deg with the sample rotation and realigning:
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.034 | V2O3 608 E8 Rest |
motor scan Y | ||
| 035 | motor scan Z | |||
| 036 | EXAFS scan E perp. c (?) 149 K, 73.8 Ohm user comments in data file wrong |
|||
| recalibrated mono from 5448 retuned optics |
||||
| 037 | dito user comments in data file wrong VERY NOISY |
|||
| recalibrated mono from 5442 | ||||
| retuned optics realigned sample |
||||
| 038 | EXAFS scan dito user comments corrected |
|||
| mono recalibrated from 5445.5 | ||||
| 039 | EXAFS scan | |||
| mono recalibrated from 5448.5 | ||||
| 040 | ??????????????? |
Lowering temp by 5 K to 144 K
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.041 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 144 K, 77.5 Ohm |
||
| recalibrated mono from 5448 retuned optics realigned sample |
||||
| 042 | EXAFS scan E perp. c (?) 144 K, 77.5 Ohm |
|||
| recalibrated mono from 5440 | ||||
| 043 | dito | |||
| recalibrated mono from 5447 |
Went to 139 K, still no sign of the MIT
| Filename | sample | comments | ranges | stepsize |
| 17Nov00.044 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 139 K, 80.5 Ohm |
||
| recalibrated mono from ??? | ||||
| 045 | dito | |||
| mono recalibrated from 5443 | ||||
| 046 | dito | |||
| 047 | dito, edge far off |
2000-11-18
Mono recalibrated from 5429 eV
Optics retuned
Sample aligned
| Filename | sample | comments | ranges | stepsize |
| 18Nov00.001 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 133 K, 91.7 Ohm |
||
| mono recalibrated from 5441 | ||||
| 002 | EXAFS scan E perp. c (?) 133 K, 92.7 Ohm |
|||
| 003 | follow up scan mono far off! |
|||
| recalibrated mono from 5432 | ||||
| 004 | 91.7 Ohm | |||
| recalibrated mono from 5442 (?) realigned sample |
||||
| 005 | 92.2 Ohm | |||
| recalibrated mono from 5445 |
Changed temp to 130 K (126.9 Ohms) - we're close to the transition now!
Changed the mono calibration by 10 eV to be on the safe side.
Taking a scan after retuning the optics, will realign the sample after this
scan
| Filename | sample | comments | regions | stepsize |
| 18Nov00.006 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 130 K, 126.9 Ohm |
||
| recalibrating mono from 5450 to 5475 (instead of 5465) retuning optics |
||||
| 007 | motor scan Z | |||
| 008 | motor scan Y | -1, 1 | .1 | |
| 009 | EXAFS scan E perp c (?) 130 K, 138.6 Ohm |
|||
| recalibrating mono from 5452 to 5475 (instead of 5465) | ||||
| 010 | 137.2 Ohm | |||
| recalibrating mono from 5456 to 5475 (instead of 5465) | ||||
| 011 | motor scan Y | -1, 1 | .1 | |
| 012 | morot scan Z | |||
| 013 | EXAFS scan 135.5 Ohm |
Looks like there was some nonlinearity in the signal: Glitches at 6000eV (11.56 inverse Angstroem) seem to be related to glitches in Io.
Finally I found the gas mix calculation window in the XDAC (its a
subfunction of the Detectors screen). Changing the gas mix from N2 to 1:1 N2:He
should reduce absorption of the Io chamber from 14 % to 7 % at the V K edge. I
set the respective flows to 20 units on the flow meters. The Io reading before
the change was 2.76, so it should go to 1.4 when it is stable. I don't know if
I can wait that long. I start another scan at the same temp.
| Filename | sample | comments | regions | stepsize |
| 18Nov00.014 | V2O3 608 E8 Rest |
motor scan Z | -1, 1 | .1 |
| 015 | motor scan Y | |||
| 016 | EXAFS scan E perp c (?) 130 K, 135.2 Ohm Io may be drifting during scan as gas mixture changes!!! |
-200, -30, 30, 16k | 2, .5, .05k | |
| 017 | dito, plus ugly region around 5558 eV. unusable |
Checking linearity w/ timescan:
| Filename | sample | comments | regions | stepsize |
| 18Nov00.018 | V2O3 608 E8 Rest |
time scan 1 Al foil does not change linearity, scan aborted after that. |
||
| 019 | motor scan Y | |||
| 020 | moto scan Z |
Going down to 127 K and realigning the sample
| Filename | sample | comments | regions | stepsize |
| 18Nov00.022 | V2O3 608 E8 Rest |
motor scan Z | ||
| 023 | motor scan Y | |||
| 024 | EXAFS scan E perp c (?) 127 K, 165.9 Ohm |
|||
| recalibrated mono realigned sample |
||||
| 025 | EXAFS scan 127 K, 159.3 Ohm probably the same glitches like in scan 026?! |
|||
| mono recalibrated | ||||
| 026 | EXAFS scan 127 K, 159.2 Ohm noticed strange dips @ 5682.5 and 6005 eV (Io related glitches) |
|||
| 027 | dito, glitches better |
Going to 124 K
| Filename | sample | comments | regions | stepsize |
| 18Nov00.028 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 124 K, 189 Ohm wrong user comments in data file, no sample alignment before scan |
||
| recalibrated mono realigned sample Found strange double peak structure in Z and Y scans. Has the sample cracked??? |
||||
| 029 | EXAFS scan 124 K, 197.8 Ohm |
|||
| 030 | 198 Ohm scan aborted by injection above 13 inverse Angstroem |
Thought over another possibility why the peaks in Io could result in peaks in I_t: Possible pinholes in the sample! So part of Io is "leaking" through the sample without attenuation and reproducing the Io signal shape!
Gone to 120 K and 283 Ohms, aligning sample after injection while beamline
is warming up:
| Filename | sample | comments | regions | stepsize |
| 18Nov00.031 | V2O3 608 E8 Rest |
motor scan Y | -2, 2 | .1 |
| 032 | motor scan Z | |||
| 033 | edge scan | |||
| 034 | EXAFS scan E perp c (?) 120 K, 283 Ohm |
|||
| mono recalibrated optics retuned |
||||
| 035 | dito | |||
| 036 | dito |
Reduced Temp. to 116 K (395 Ohms)
| Filename | sample | comments | regions | stepsize |
| 18Nov00.037 | V2O3 608 E8 Rest |
motor scan Y | -1, 1 | .1 |
| 038 | motor scan Z | |||
| 039 | EXAFS scan E perp c (?) 116 K, 393 Ohm |
|||
| 040 | 407 Ohm |
Going down to 112 K (503 Ohms):
| Filename | sample | comments | regions | stepsize |
| 18Nov00.041 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 112 K, 503 Ohm |
||
| 042 | dito |
2000-11-19
Still there on a new day!
Going to 105 K
| Filename | sample | comments | regions | stepsize |
| 19Nov00.001 | motor scan Y | |||
| 19Nov00.002 | motor scan Z | |||
| no change in sample positions! | ||||
| 19Nov00.003 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 105 K, 730 Ohm |
||
| 004 | dito, 780 Ohm |
Skipping 95 K and 1265 Ohm and going to 90 K and 1665 Ohms right away.
| Filename | sample | comments | regions | stepsize |
| 19Nov00.005 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 90 K, 1670 Ohm Io related glitch at 5679.6 eV looks bumpy overall |
||
| realigned sample, retuned optics (no longer changing!) and recalibrated mono | ||||
| 006 | 1785 Ohm smoother! |
|||
| 007 | 1830 Ohm |
I try to go down to 30 K. The temperature is reached for a few moments but going up again, then. The part of the cup that covers the 2nd stage cooling is icing. Probably the Al foil I used for insulating the 2nd stage touches the cup (got loose?). Shit! But I haven't found Johnny's super-insulation foil that worked so well the last times!
Spent a great hour from 4 a.m. to 5 a.m warming up the Displex, changing the
Al foil that made thermal contact to the cup and cooling the Displex down
again.
Sample alignment fortunately still ok, just doing the usual fine tuning:
| Filename | sample | comments | regions | stepsize |
| 19Nov00.008 | V2O3 608 E8 Rest |
motor scan Y | -2, 2 | .1 |
| 009 | motor scan Z | |||
| 010 | EXAFS scan E perp c (?) 30 K, 24 kOhm |
|||
| 011 | 30 kOhm | |||
| 012 | 29.5 kOhm |
Heating to 150 K and taking a last scan
| Filename | sample | comments | regions | stepsize |
| 19Nov00.013 | V2O3 608 E8 Rest |
EXAFS scan E perp. c (?) 150 K, 85 Ohm |
Returned at lunchtime after 5 hours of sleep and really feeling better now.
Displex had warmed up. We rotated the sample to the second orientation which
should be E parallel c.
Unfortunately one of the wires for the online resistance measurement went off
together with a part of the glass the sample is mounte on. So no more
resistance values for the new temperature series.
While I brought Vitaly to the train and picked up Scott, The Displex cooled
down to 150 K and the beamline warmed up (hopefully). Aligning the sample and
taking the first EXAFS scan:
| Filename | sample | comments | regions | stepsize |
| 19Nov00.014 | V2O3 608 E8 Rest |
motor scan Y | ||
| 015 | motor scan Z | |||
| 016 | EXAFS scan E par. c (?) 150 K |
|||
| 017 | dito | |||
| retuning optics, realigning sample | ||||
| 018 | dito Glitch at 5679.6 and 5901 |
going to 133 K, retuning optics (didn't change) and realigning sample
| Filename | sample | comments | regions | stepsize |
| 19Nov00.019 | V2O3 608 E8 Rest |
motor scan Y | ||
| 020 | motor scan Z | |||
| 021 | EXAFS scan E par. c (?) 133 K |
|||
| 022 | dito |
going to 130 K, retuning optics (didn't change) and realigning sample
| Filename | sample | comments | regions | stepsize |
| 19Nov00.023 | V2O3 608 E8 Rest |
motor scan Y | ||
| 024 | motor scan Z | |||
| 025 | EXAFS scan E par. c (?) 130 K |
|||
| 026 | dito |
going to 127 K and realigning sample
| Filename | sample | comments | regions | stepsize |
| 19Nov00.027 | V2O3 608 E8 Rest |
motor scan Y | ||
| 028 | motor scan Z | |||
| 029 | EXAFS scan E par. c (?) 127 K |
|||
| injection @19.00 | ||||
| 030 | motor scan Y | |||
| 031 | motor scan Z | |||
| 032 | EXAFS scan E par c (?) 127 K ugly Io due to moving source after injection. Probably scan is unusable. |
|||
| aligning sample | ||||
| 033 | dito, looks noisy, too | |||
| aligned sample | ||||
| 034 | dito |
calibration: two subpeaks in calibration region; chose higher one.
Recalibrated 5454.5 to 5475.
Aligned mirrors.
Set to 124K (dropped to 123.7 while setting)
| Filename | sample | comments | regions | stepsize |
| 19Nov00.035 | EXAFS scan E par c (?) 124 K |
|||
| Recalibrated 5447 to 5775 Realigned mirrors, but they seem to be stabilizing |
||||
| 036 | dito, unusably noisy | |||
| realigned sample | ||||
| 037 | dito, looky much better at the first glance, but not at second | |||
| 038 | next try, worse again. Killed somewhere in the middle | |||
| aligning sample: | ||||
| 039 | motor scan Y | |||
| 040 | motor scan Z | |||
| 041 | EXAFS scan, again looks much better, but different than all the previous ones. So I hav to take another one: |
|||
| 042 | EXAFS scan, 124 K |
2000-11-20
Still there on a new day!
Finally got two scans for 124 K that look alike and finishing that temperature
that really gave me a hard time!
Going to 120 K
| Filename | sample | comments | regions | stepsize |
| 20Nov00.001 | V2O3 608 E8 Rest |
motor scan Y | ||
| 002 | motor scan Z | |||
| 003 | EXAFS scan E par. c (?) 120 K |
|||
| realigning sample | ||||
| 004 | dito Glitch at 6000.5eV |
Strange things happen: Both scans are usable!
Changing temp to 116 K, doing the usual tuning and alignment:
| Filename | sample | comments | regions | stepsize |
| 20Nov00.005 | V2O3 608 E8 Rest |
motor scan Y | ||
| 006 | motor scan Z | |||
| 007 | EXAFS scan E par. c (?) 116 K |
|||
| sample realigned | ||||
| 008 | dito Glitch at 5676.8? |
|||
| retuning optics, realigning sample | ||||
| 009 | dito | |||
| 010 | 2nd scan in set without recalibration |
Next morning:
Went to the beamline befor breakfast to warm up the mirrors and change the
temperature of the sample to 112 K
| Filename | sample | comments | regions | stepsize |
| 20Nov00.011 | V2O3 608 E8 Rest |
motor scan Y | -2, 2 | .1 |
| 012 | motor scan Z | |||
| 013 | EXAFS scan E par. c (?) 112 K |
|||
| realigned sample | ||||
| 014 | dito | |||
| 015 | motor scan Y | |||
| 016 | motor scan Z | |||
| 017 | EXAFS scan E par. c (?) 112 K looks glitchy, but isn't |
|||
| 018 | dito glitches at 5623 and 5682?! |
Once it's working, it's really great: got two identical scans up to 14 A^-1
!
Going to 90 K now:
| Filename | sample | comments | regions | stepsize |
| 20Nov00.019 | V2O3 608 E8 Rest |
motor scan Y | ||
| 020 | motor scan Z | |||
| 021 | EXAFS scan E par c (?) 90 K looks good, but large E shift |
|||
| checked alignment of sample. Did not move | ||||
| 022 | dito | |||
| 023 | dito |
Going down to 30 K
| Filename | sample | comments | regions | stepsize |
| 20Nov00.024 | V2O3 608 E8 Rest |
motor scan Y | ||
| 025 | motor scan Z | |||
| 026 | EXAFS scan E par. c (?) 30 K Glitch at 5991.5 |
|||
| checking alignment (didn't change) | ||||
| 027 | dito | |||
| realigning sample | ||||
| 028 | dito Glitch at 5676.6 |
Finally heating to 150 K again:
| Filename | sample | comments | regions | stepsize |
| 20Nov00.029 | V2O3 608 E8 Rest |
motor scan Y | ||
| 030 | motor scan Z | |||
| 031 | EXAFS scan E par. c (?) 150 K |
FINISHED WITH V2O3 FOR THIS RUN!!!
Checking energy calibration of monochromator now, using Ti, V and Cr foil:
| Filename | sample | comments | regions | stepsize |
| 20Nov00.032 | Ti foil V foil Cr foil |
edge scans | -700, -550, -450, -50, 50, 475, 575 | 5, .5, 5, .5, 5, .5 |
| 033 | V foil no Cr foil Mn fiol |
edge scan V removed Cr instead of Ti foil, $%&*+!§ |
-100, 100 | 2 |
| 034 | scanning all three edges well, no Cr edge appeared ... |
-50, 50, 475, 575, 1025, 1125 | .5, 5, .5, 5, .5 | |
| 035 | V, Cr, Mn foils (really this time!) |
V edge scan | as 033 | |
| 036 | all three edges now | as 034 |
| edge | recorded position | expected position | drift |
| V | 5462 | 5465 | -3 |
| Cr | 5970.5 | 5989 | -18.5 / 524eV |
| Mn | 6499.5 | 6539 | -39.5 / 550+524eV |
| Filename | sample | comments | regions | stepsize |
| 20Nov00.037 | V, Cr, Mn foil | V edge scan edge was at 5445 (?); calibrated it to 5465 |
-100, 100 | 2 |
| 038 | all three edges | -50, 50, 475, 575, 1025, 1125 | 1, 10, 1, 10, 1 |
| edge | recorded position | expected position | drift |
| V | 5463 | 5465 | -2 |
| Cr | 5969 | 5989 | -20 / 524eV |
| Mn | 6500 | 6539 | -39 / 550+524eV |
| Filename | sample | comments | regions | stepsize |
| 20Nov00.039 | V, Cr, Mn foil | V edge scan edge was at 5449; calibrated it to 5465 |
-100, 100 | 2 |
| 040 | EXAFS scan similar to the ones for V2O3 | -200, -30, 30, 17k | 2, .5, .05k |
| edge | recorded position | expected position | drift |
| V | 5459.5 | 5465 | -5.5 |
| Cr | 5965 | 5989 | -24 / 524eV |
| Mn | 6500 | 6539 | -39 / 550+524eV |
Putting in the Mn foil.
| Filename | sample | comments | regions | stepsize |
| 20Nov00.041 | Mn foil | motor scan Z at 300 eV above Mn edge | -4, 4 | .1 |
| 042 | Y | .2 | ||
| decreased gain on I_t to 1e8 | ||||
| 043 | edge scan | -100, 100 | 2 | |
| 044 | Mn foil Co reference |
EXAFS scan | -200, -30, 30, 18k | 2, .5, .05k |
| 045 | garbage | |||
| 046 | checking reference detector | 1120, 1220 | 5 | |
| 047 | 2 | |||
| 048 | turned I_ref "chamber" around | |||
| 049 | Mn foil reference detector w/out foil |
|||
| 050 | Mn foil and Co foil between Io and I_t detectors | reference signal without foil | ||
| 051 | 1020, 1220 | |||
| found the Co edge 170 eV off and with an edge jump of 4! | ||||
| 052 | quick scan through whole region | -200, 17.5k | 5 | |
| edge was too high, this time --> recalibrated | ||||
| 053 | EXAFS scan | -200, -30, 30, 18k | 2, .5, .05k | |
| 054 | Co foil only | Scan as planned for Mn foil to check if Co signal is really smooth in th
desired energy range using integration time 1 for this scan to speed things up LOST LOCK!!! |
-200, -50, 50, 16.5k, 18k | 5, 1, .05k, 2 |
| 055 | dito LOST LOCK!!! |
|||
| 056 | Co edge scan | 1020, 1220 | 5 | |
| 057 | like 054 | 5,5,5,5 | ||
| 058 | garbage |
2000-11-21
| 21Nov00.001 | Mn foil and Co foil | starting real (!) EXAFS scan on Mn now LOST LOCK!!! |
-200, -50, 50, 16.5k, 18k | 5, 1, .05k, 2 |
| retuned mirror 2 at 450eV above edge increased sensitivity on lockin to 500mV from 1V hopefully that helps and doe not make things worse! |
||||
| 002 | another EXAFS scan LOST LOCK!!! |
dito | dito | |
| Adjusted mirrors at 800 eV above edge. "Lowered" sensitivity to 100 mV. Checked by moving monochromator directly up and down...seems OK! |
||||
| 003 | another EXAFS scan worked--yay! |
-200,-50,50, 16.5k,18k |
5,1,.05k,2 | |
| 004- | more EXAFS scans edge far off! |
-200,-50,50, 16.5k,18k |
5,1,.05k,2 |
returned next morning:
| Filename | sample | comments | regions | stepsize |
| 21Nov00.006 | Mn foil + Co foil | edge scan | -200, 0 | ? |
| recalibrated Mn edge from 6434 eV | ||||
| 007 | EXAFS scan nearly lost lock, but recovered after one point |
-200, -50, 50, 16.5k, 18k | 5, 1, .05k, 2 | |
| reaclibrated Mn edge (was approx. 20eV low) | ||||
| trying a even higher sensitivity of 50mV (which corresponds to range (V/div) on oszi) | ||||
| optimizing the flux 800 eV above the edge leads to a very low Io at the begin of the scan (20% of the value at +800eV). I'm trying +600eV. Does not change anything. | ||||
| set bias to 0 at -200eV | ||||
| 008 | new EXAFS scan at 250K have been moving back and forth ca couple of times: edge is ~40eV too low. LOST LOCK!!! |
|||
| returning to sensitivity of 100mV | ||||
| 009 | EXAFS scan, again. LOST LOCK |
|||
| as "rocking" sine curve was bumping around at high energies, I changed the gain a bit. the old position is marked with electrical tape (turn counterclockwise to return). Trying another scan | ||||
| 010 | EXAFS scan LOST LOCK AGAIN. |
|||
| recalibrated edge (was at 6484) | ||||
| 011 | quick scan with integration times of 1 sec to check it again. Did not set
bias to 0 at begin of scan this time. Went through. |
|||
| changed gain a bit, retuned optics at +700eV and set bias to 0 there | ||||
| 012 | doing an EXAFS scan with 2 sec integration time. Worked! |
|||
| Probably setting the bias at 700eV above the edge (and not at the beginning of the scan) is the clue??? |
Reducing temp to 200 K.
| Filename | sample | comments | regions | stepsize |
| 21Nov00.013 | Mn foil + Co foil | EXAFS scan 200 K |
||
| 014 | dito |
Reducing temp to 150 K.
| Filename | sample | comments | regions | stepsize |
| 21Nov00.015 | Mn foil + Co foil | EXAFS scan 150 K |
||
| 016 | dito |
Reducing temp to 100 K.
| Filename | sample | comments | regions | stepsize |
| 21Nov00.017 | Mn foil + Co foil | EXAFS scan 100 K |
||
| 018 | garbage? | |||
| 019 | 100 K EXAFS scan |
Reducing temp to 50 K.
| Filename | sample | comments | regions | stepsize |
| 21Nov00.020 | Mn foil + Co foil | EXAFS scan 50 K |
||
| 50 K EXAFS scan |
Returned from dinner, put in Cr foil and Fe as a reference (16.7A^-1 above
Cr edge).
Tried to tune optics. Flux was very low, so I tried to tune it for the V K edge
where I knew the flux had to be around 1.0; was .01 what did not add to my
pleasure. Noticed we had run out of He. Probably that already caused the flux
to be so low during the Mn scans. Got a new bottle of compressed He for 34.91$
from the Stockroom.
Flux is slowly increasing.
Changing gain to 500mV
Starting 1st scan while gas flow still is settling
| Filename | sample | comments | regions | stepsize |
| 21Nov00.022 | Cr foil + Fe foil | EXAFS scan 300 K 1 sec integration times |
-200, -50, 50, 16.0k, 17.5 k | 5, 1, .05k, 2 |
| 023 | EXAFS scan 300 K 2 sec integration |
|||
| 024 | dito LOST LOCK! |
|||
| changing sensitivity to 1V (lockin is in overload) | ||||
| 025 | new EXAFS scan 300 K |
Lowering temperature to 250 K.
| Filename | sample | comments | regions | stepsize |
| 22Nov00.001 | Cr foil + Fe foil | EXAFS scan | -200,-50,50,16.0k,17.5k | 5,1,.05k,2 |
| 002 | dito Glitch in Io at 6967 |
|||
| adjusted bias at "edge" (actually above because of drift) Moved 50 below the edge and recalibrated as the edge |
||||
| 003 | Cr foil + Fe foil | 200 K EXAFS scan | -200,-50,50,16.0k,17.5k | 5,1,.05k,2 |
| 004 | ditto Before scanning, moved 20 eV below edge and recalibrated. |
|||
| 005 | Cr foil + Fe foil | 150 K EXAFS scan | -200,-50,50,16.0k,17.5k | 5,1,.05k,2 |
| 006 | trash | |||
| 007 | trash | |||
| recording last scan at 150 K the next morning: | ||||
| 008 | 150 K EXAFS scan I_t detector may be saturated or nonlinear responding (went to ~11V, full saturation occurs at ~12V) above 6900eV (~15A^-1) |
Starting 100 K scans:
| Filename | sample | comments | regions | stepsize |
| 22Nov00.009+010 | Cr foil + Fe foil | EXAFS scans at 100 K #010 lost lock at 6964 |
||
| 011 | dito |
Starting 50 K scans:
| Filename | sample | comments | regions | stepsize |
| 22Nov00.012 | Cr foil + Fe foil | EXAFS scans at 050 K wrong user comments in data file (still says 100 K) |
||
| 013 | dito |
| Filename | sample | comments | regions | stepsize |
| 22Nov00.014 | _NO_ Cr foil + Fe foil | EXAFS scan on Cr edge with Fe foil only to chekc smoothness of Fe foil in
Cr region integration times set to 1 sec |
-200, -50,50, 16k, 17.5k | 5,2,.05k,2 |