NSLS - X23B - 11-2000

2000-11-14
Arrived with Jihane. We mounted the Displex, connected the cooling water and wired it up.

2000-11-15
Got SAF form and enabled beamline.
Johnny is using a new pumping stand for the mono chamber (pair of Si crystals). The old pump caused to much vibrations. As the pumping stand is located so that it can be kicked out by X23A users, checking the mono tank vacuum would be pretty wise.

There are a few important changes in the handling of the lockin:
On starting XDAC turn on Digital Feedback in the Mono setup menu.
Changing the gain is not necessary any more. (That is what the feedback does.)
Adjust mirrors in the middle of the scan, then return to the beginning and set BIAS to zero there.

We enabled the beamline and let it warm up.
A beamdump occurs just before I wanted to start the first edge scan on a V foil. They had to access the ring tunnel and repair a cooling water supply.
After the injection (15:30) I start the edge scan:
 

Filename sample comments ranges stepsize # scans
15Nov00.001 V foil edge scan -100, 100 2 1
calibrated mono to V edge
15Nov00.002 V foil near edge scan -30, 30 1 1
calibrated mono to V edge

Another beamdump at 16:30!

I checked which motor numbers are used:
The correct names are Sample-Horiz, Sample-Vert and Sample-Rot with the numbers 3,4,1

mounted sample 1156 K2 S...  with E vector parallel c axis
 

Filename sample comments ranges stepsize # scans
15Nov00.003 V2O3
1156 K2
motor scan Z
(wrong slice number in data comments)
-4, 4 .1 1
calibrated motor
15Nov00.004 motor scan Y -2, 2 .1 1
calibrated motor
15Nov00.005 motor scan Z -2, 2 .1 1
recalibrated motor
15Nov00.006 edge scan
Gain wrong on Io, has to be 1e7
-100, 100 1 1

After the third beamdump for this afternoon, I went to pick up Mihail and Roman from the station.
The beam returned at 8.30 p.m. We let the mirrors warm up again and started with a readjustment of the sample:
 

Filename sample comments ranges stepsize # scans
15Nov00.007 V2O3
1156 K2
motor scan Z -2, 2 .1 1
calibrated motor
15Nov00.008 motor scan Y -2, 2 .1 1
calibrated motor
15Nov00.009 edge scan -100, 100 1 1
recalibrated mono
15Nov00.010 EXAFS scan
Gain wrong on Io, has to be 1e7
-200, -10, 30, 16k 2,2,2 2, .5, .05k
mono recalibrated (from 4551eV)
15Nov00.011
15Nov00.012
Gain wrong on Io, has to be 1e7

2000-11-16
Johnny obvoiusly was here lat at night (11 p.m.?). He left me a note, telling that he changed the scale on the oszi and turned on Io plotting in the XDAC.
At the beginning of the third scan it lost lock and Johnny retuned the optics and started the scans over.
He reduced sensitivity from 500mV to 1V on the lockin amp. as is was in overload.

He also mentioned that Io looked very big to him and suggested that I should add som He to Io or to shut down the slits a bit . Otherwise Io might not be linear.
Gain on Io is quite high (1e9), so probably thats not necessary. I have to talk to Johnny about that today. Closing the slits should not hurt in any case with my sample size.
 
 

Filename sample comments ranges stepsize
15Nov00.013 - 017 V2O3
1156 K2
EXAFS scans
#017 seems to be empty
Gain wrong on Io, has to be 1e7
-200, -10, 30, 16k 2, .5, .05k

Emergency ring access seems to take longer than they thought. They're starting the injection at 8:45.

At some point I changed the Gain reading for Io in the XDAC to its correct value of 1e7
 

Filename sample comments ranges stepsize
16Nov00.001 V2O3
1156 K2
time scan 2 check lin. of Io:
1st part: no Al foil
2nd part: 1 Al foil
3rd part: 2 Al foils
0, 1000 1

Noticed a hole in the aluminized mylar foil at the end of the beamline extension in the hutch. fixed it.
Reduced exit slits to 1 x 1 mm.
I_t is significantly reduced. Doing motor scans again:
 

Filename sample comments ranges stepsize
16Nov00.002 V2O3
1156 K2
motor scan Z
no change in Z position
wrong user comments in data file
-2, 2 .1
16Nov00.003 motor scan Y
only small change
wrong user comments in data file
004 time scan again 0, 1000 1

No good result: linearity changes 5-10%, Johnny says tey got it to as good as .5%.
So I open up the slits again, increasing the I_t.
Using mixture of N2 and He for Io chamber now. Flowmeter shows 40 for N2 and 10 for He. Io reduced from 3.05 to 2.65. Testing linearity again:
 

Filename sample comments ranges stepsize
16Nov00.005 V2O3
1156 K2
time scan 0, 1000 1
For some reason, the Io/I_t ration does not stay constant this time, but its increasing with time (also without changing the intensity with Al foil). Perhaps the gas flows need some time to stabilize.

A propos gas flows:
 

Device He N2 Ar
Beamline window red  17 - - - -
Beamline extension green  90 - - - -
Io (still testing) blue 30 white 10 - -
I_t - - gray 40 - -
I_Lytle (not in use) - - - - yellow (33) 0
(not in use) brown
(not in use) black

Yesterday we connected the wrong hoses and were blowing N2 into the hutch instead of into the ion chambers. That problem is fixed, now.
 

Filename sample comments ranges stepsize
16Nov00.006 V2O3
1156 K2
time scan
lost beam in the region of 200 sec
0,1,2 layers of Al foil
0, 1000 1
007 check for stabilizing Io
changed He/N2 to 20/20 at 144
008 continuing from scan 007, without glitch
009 another time scan
changed He/N2 to 25/15 at 80
010 dito, 009 was slowly stabilizing at 1000

I decide to leave the gas flows constant now. They are:
 

Device He N2 Ar
Beamline window red  17 - - - -
Beamline extension green  90 - - - -
Io (still testing) blue 25 white 15 - -
I_t - - gray 40 - -
I_Lytle (not in use) - - - - yellow (33) 0
(not in use) brown
(not in use) black

Doing another linearity check. Hopefully it's the last one:
 

Filename sample comments ranges stepsize
16Nov00.011 V2O3
1156 K2
time scan 4 linearity check 0, 1000 1
No linearity at all. Stopping He flow and just using N2 again.
 
Filename sample comments ranges stepsize
16Nov00.012 V2O3
1156 K2
time scan 4 linearity check
2 foils,
1 foil,
no foil
0, 1000 1
Now, although the ratio is still changing with time (takes some time until He is blown out of the chamber),  changing the number of foils dosn't seem to have any effect.

Final gas flow table:
 

Device He N2 Ar
Beamline window red  17 - - - -
Beamline extension green  90 - - - -
(not in use) blue 0
Io - - white 40 - -
I_t - - gray 40 - -
I_Lytle (not in use) - - - - yellow (33) 0
(not in use) brown
(not in use) black

During the overnight scans, the energy has shifted by MORE THAN 100 eV. I corrected that this morning, but rather do another edge scan:
 

Filename sample comments ranges stepsize
16Nov00.013 V2O3
1156 K2
edge scan -100, 100 1
recalibrated mono

Edge was off  50 eV
 

Filename sample comments ranges stepsize
16Nov00.014 V2O3
1156 K2
motor scan Z -2, 2 .1
015 motor scan Y
016 EXAFS scan
aborted after edge: was shifted over 10 eV
-200, -10, 30, 16k 2, .5, .05k
017 EXAFS scan -200, -30, 30, 16k 2, .5, .05k

Switched sample to have E perpendicular c now. Doing motor scans
 

Filename sample comments ranges stepsize
16Nov00.018 V2O3
1156 K2
E perp c now
motor scan Z
-4, 4 .1
019 motor scan Y
did a burnpaper check and moved stage horizontally
020 motor scan Z
021 motor scan Y
022 EXAFS scan
E perp c
killed by PP after edge
-200, -30, 30, 16k 2, .5, .05k
mono recalibrated
023 EXAFS scan
E perp c
MONO RECALIBRATED
024 dito
MONO RECALIBRATED
025 dito

Mounted new sample 608 E8 Rest w/ E perpendicular c (??? Sample orientation has to be rechecked, Perpendicular orientation results from comparison with spectra of other samples). Using graphite for thermal contact and wiring for resistance measurement.

Starting with motor scans:
 

Filename sample comments ranges stepsize
16Nov00.026 V2O3
608 E8 R
motor scan Z -2, 2 .1
027 motor scan Y
028 edge scan -100, 100 1
RECALIBRATED MONO
029 EXAFS scan
Obviously E perp. c
-200, -30, 30, 16k 2, .5, .05k
030 EXAFS scan
Obviously E perp. c

After dinner we decide to go on with Mike's and Roman's experiment.
We tilt the Displex out of the way and put up their Ti foil.
 

Filename sample comments ranges stepsize
16Nov00.031 Ti foil edge scan
no current
-100, 100 2
recalibrated mono
16Nov00.032 near edge scan
no battery
-30, 30 0.3
16Nov00.033 near edge scan
1 battery connected
034 without battery again
035 with battery
036 with 4 batteries

 
 

2000-11-17
Just looked up the transition temperature of charge 608: 141 K on cooling, 148 K on warming up.
We have set up the Displex. I took my chance and spilled some cooling water on Roman :-(
The forepump runs quite smoothly and does not seem likely to fall off the stool.
Resistance of sample is 76 Ohms @ 220 K

Looking for sample with motor scans:
 

Filename sample comments ranges stepsize
17Nov00.001 V2O3
608 E8 Rest
motor scan Z -2, 2 .1
002, 003, 004 motor scans Y different ranges different stepsizes
005 motor scan Z -4, 4 .2
006 motor scan Y -2, -2 .1
007 motor scan Z
008 edge scan -100, -30, 30, 100 2,1,2

Found the sample, calibrated the edge, retuned the optics.
Starting an EXAFS scan now
 

Filename sample comments ranges stepsize
17Nov00.009 V2O3
608 E8 Rest
EXAFS scan
E perp c (?)
200 K, 81.4 Ohm
-200, -30, 30, 16k 2, .5, .05k
mono recalibrated (5452 -> 5465)
010 200 K, 55 Ohm (new battery in Multimeter)

Johnny tried something on the mono motors which did not work yet. He restored the old settings and I take another scan after retuning the optics.
 

Filename sample comments ranges stepsize
17Nov00.011 V2O3
608 E8 Rest
EXAFS scan
E perp c (?)
200 K, 55.9 Ohm
ABORTED
-200, -30, 30, 16k 2, .5, .05k

In each of the scans there are strange structures in the preedge region ( around 5300 eV) that change slightly from scan to scan. But that was not the reason why I aborted the scan. I noticed that the sample position has changed and the I_t signal was very low.
So doing motor scans again:
 

Filename sample comments ranges stepsize
17Nov00.012 V2O3
608 E8 Rest
motor scan Z -2, 2 .1
013 motor scan Y
014 edge scan
Changed gain on I_t to 1e10, but forgot to measure offsets
-50, 50 1
calibrated mono (5432 -> 5465)

Measured offsets now, after changing gain on I_t to 1e10. Starting an EXAFS scan:
 

Filename sample comments ranges stepsize
17Nov00.015 V2O3
608 E8 Rest
EXAFS scan
E perp c (?)
200 K, 56.1 Ohm
I_t detector went to nearly 11 V at the very end of the scan.
-200, -30, 30, 16k 2, .5, .05k
016 testing alternative mono drive (energy driving wrong way)
017 dito
018 edge scan
170 K, 64.1 Ohm
-100, 100 2

Reset I_t gain to 1e9 and measured offsets (really did it this time befor the first scan!).
Probably low gain is better for the scan, even if signal is low. But Johnny says that linearity is not guaranteed abov 8 Volts - and I was reaching 11 V!
 

Filename sample comments ranges stepsize
17Nov00.019 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
155K, 70 Ohm
-200, -30, 30, 16k 2, .5, .05k
calibrated mono (5453 -> 5465)
020 motor scan Y
021 motor scan Z
022 EXAFS scan
E perp. c (?)
155K, 70.5 Ohm
mono recalibrated ( 5447 -> 5465 )
023 EXAFS scan
E perp. c (?)
155K, 70.1 Ohm
024 dito
mono recalibrated (5429 -> 5465)

Went to 149 K (73 Ohm), maybe the transition occurs at a temp. a bit lower than measured in Augsburg.
Beam dumped (2:45). I "stand by".

New injection, new photons, new alignment:
 

Filename sample comments ranges stepsize
17Nov00.025 V2O3
608 E8 Rest
motor scan Z -2,2  .1
026 motor scan Y

Just talked to Joe. He will have some beamtime in Feb. or March.
He looked at my k-space data of the 155 K spectra and was suspicious about the sharp peak at 12 A^-1. I will rotate the sample a few degrees after the next scan and see if that changes (and is a Bragg peak then) or if it's really in the data.
 

Filename sample comments ranges stepsize
17Nov00.027 V2O3
608 E8 Rest
EXAFS scan
E perp c (?)
149 K, 73.5 Ohm
-200, -30, 30, 16k 2, .5, .05k
recalibrated mono from 5440
rotated sample to 4 units of "sample rot" motor (should be 4 or 8 deg.), was at 0 units before moving.
028 motor scan Y
moved to -.3 but did not recalibrate motor
029 motor scan Z
moved to +.3 but did not recalibrate motor
030 EXAFS scan
149 K, 73.7 Ohm
recalibrated mono from 5447
rotated sample to -4 units
031 motor scan Z
moved to .4 w/out calibration
032 motor scan Y
moved to -.2 w/out calibration
033 EXAFS scan
dito
recalibrated mono from 5450

I believe that the sharp structure is just noisy. All three scans look quite the same.
Returning to 0 deg with the sample rotation and realigning:
 

Filename sample comments ranges stepsize
17Nov00.034 V2O3
608 E8 Rest
motor scan Y
035 motor scan Z
036 EXAFS scan
E perp. c (?)
149 K, 73.8 Ohm
user comments in data file wrong
recalibrated mono from 5448
retuned optics
037 dito
user comments in data file wrong
VERY NOISY
recalibrated mono from 5442
retuned optics
realigned sample
038 EXAFS scan
dito
user comments corrected
mono recalibrated from 5445.5
039 EXAFS scan
mono recalibrated from 5448.5
040 ???????????????

Lowering temp by 5 K to 144 K
 

Filename sample comments ranges stepsize
17Nov00.041 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
144 K, 77.5 Ohm
recalibrated mono from 5448
retuned optics
realigned sample
042 EXAFS scan
E perp. c (?)
144 K, 77.5 Ohm
recalibrated mono from 5440
043 dito
recalibrated mono from 5447

Went to 139 K, still no sign of the MIT
 

Filename sample comments ranges stepsize
17Nov00.044 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
139 K, 80.5 Ohm
recalibrated mono from ???
045 dito
mono recalibrated from 5443
046 dito
047 dito, edge far off

 
 

2000-11-18

Mono recalibrated from 5429 eV
Optics retuned
Sample aligned
 

Filename sample comments ranges stepsize
18Nov00.001 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
133 K, 91.7 Ohm
mono recalibrated from 5441
002 EXAFS scan
E perp. c (?)
133 K, 92.7 Ohm
003 follow up scan 
mono far off!
recalibrated mono from 5432
004 91.7 Ohm
recalibrated mono from 5442 (?)
realigned sample
005 92.2 Ohm
recalibrated mono from 5445

Changed temp to 130 K (126.9 Ohms) - we're close to the transition now!
Changed the mono calibration by 10 eV to be on the safe side.
Taking a scan after retuning the optics, will realign the sample after this scan
 

Filename sample comments regions stepsize
18Nov00.006 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
130 K, 126.9 Ohm
recalibrating mono from 5450 to 5475 (instead of 5465)
retuning optics
007 motor scan Z
008 motor scan Y -1, 1 .1
009 EXAFS scan
E perp c (?)
130 K, 138.6 Ohm
recalibrating mono from 5452 to 5475 (instead of 5465)
010 137.2 Ohm
recalibrating mono from 5456 to 5475 (instead of 5465)
011 motor scan Y -1, 1 .1
012 morot scan Z
013 EXAFS scan
135.5 Ohm

Looks like there was some nonlinearity in the signal: Glitches at 6000eV (11.56 inverse Angstroem) seem to be related to glitches in Io.

Finally I found the gas mix calculation window in the XDAC (its a subfunction of the Detectors screen). Changing the gas mix from N2 to 1:1 N2:He should reduce absorption of the Io chamber from 14 % to 7 % at the V K edge. I set the respective flows to 20 units on the flow meters. The Io reading before the change was 2.76, so it should go to 1.4 when it is stable. I don't know if I can wait that long. I start another scan at the same temp.
 

Filename sample comments regions stepsize
18Nov00.014 V2O3
608 E8 Rest
motor scan Z -1, 1 .1
015 motor scan Y
016 EXAFS scan
E perp c (?)
130 K, 135.2 Ohm
Io may be drifting during scan as gas mixture changes!!!
-200, -30, 30, 16k 2, .5, .05k
017 dito, plus ugly region around 5558 eV. unusable

Checking linearity w/ timescan:
 

Filename sample comments regions stepsize
18Nov00.018 V2O3
608 E8 Rest
time scan
1 Al foil does not change linearity, scan aborted after that.
019 motor scan Y
020 moto scan Z

Going down to 127 K and realigning the sample
 

Filename sample comments regions stepsize
18Nov00.022 V2O3
608 E8 Rest
motor scan Z
023 motor scan Y
024 EXAFS scan
E perp c (?)
127 K, 165.9 Ohm
recalibrated mono
realigned sample
025 EXAFS scan
127 K, 159.3 Ohm
probably the same glitches like in scan 026?!
mono recalibrated
026 EXAFS scan
127 K, 159.2 Ohm
noticed strange dips @ 5682.5 and 6005 eV (Io related glitches)
027 dito, glitches better

Going to 124 K
 

Filename sample comments regions stepsize
18Nov00.028 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
124 K, 189 Ohm
wrong user comments in data file, no sample alignment before scan
recalibrated mono
realigned sample
Found strange double peak structure in Z and Y scans. Has the sample cracked???
029 EXAFS scan
124 K, 197.8 Ohm
030 198 Ohm
scan aborted by injection above 13 inverse Angstroem

Thought over another possibility why the peaks in Io could result in peaks in I_t: Possible pinholes in the sample! So part of Io is "leaking" through the sample without attenuation and reproducing the Io signal shape!

Gone to 120 K and 283 Ohms, aligning sample after injection while beamline is warming up:
 

Filename sample comments regions stepsize
18Nov00.031 V2O3
608 E8 Rest
motor scan Y -2, 2 .1
032 motor scan Z
033 edge scan
034 EXAFS scan
E perp c (?)
120 K, 283 Ohm
mono recalibrated
optics retuned
035 dito
036 dito

Reduced Temp. to 116 K (395 Ohms)
 

Filename sample comments regions stepsize
18Nov00.037 V2O3
608 E8 Rest
motor scan Y -1, 1 .1
038 motor scan Z
039 EXAFS scan
E perp c (?)
116 K, 393 Ohm
040 407 Ohm

Going down to 112 K (503 Ohms):
 

Filename sample comments regions stepsize
18Nov00.041 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
112 K, 503 Ohm
042 dito

 

2000-11-19

Still there on a new day!
Going to 105 K
 

Filename sample comments regions stepsize
19Nov00.001 motor scan Y
19Nov00.002 motor scan Z
no change in sample positions!
19Nov00.003 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
105 K, 730 Ohm
004 dito, 780 Ohm

Skipping 95 K and 1265 Ohm and going to 90 K and 1665 Ohms right away.
 

Filename sample comments regions stepsize
19Nov00.005 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
90 K, 1670 Ohm
Io related glitch at 5679.6 eV    looks bumpy overall
realigned sample, retuned optics (no longer changing!) and recalibrated mono
006 1785 Ohm
smoother!
007 1830 Ohm

I try to go down to 30 K. The temperature is reached for a few moments but going up again, then. The part of the cup that covers the 2nd stage cooling is icing. Probably the Al foil I used for insulating the 2nd stage touches the cup (got loose?). Shit! But I haven't found Johnny's super-insulation foil that worked so well the last times!

Spent a great hour from 4 a.m. to 5 a.m warming up the Displex, changing the Al foil that made thermal contact to the cup and cooling the Displex down again.
Sample alignment fortunately still ok, just doing the usual fine tuning:
 

Filename sample comments regions stepsize
19Nov00.008 V2O3
608 E8 Rest
motor scan Y -2, 2 .1
009 motor scan Z
010 EXAFS scan
E perp c (?)
30 K, 24 kOhm
011 30 kOhm
012 29.5 kOhm

Heating to 150 K and taking a last scan
 

Filename sample comments regions stepsize
19Nov00.013 V2O3
608 E8 Rest
EXAFS scan
E perp. c (?)
150 K, 85 Ohm

Returned at lunchtime after 5 hours of sleep and really feeling better now.
Displex had warmed up. We rotated the sample to the second orientation which should be E parallel c.
Unfortunately one of the wires for the online resistance measurement went off together with a part of the glass the sample is mounte on. So no more resistance values for the new temperature series.

While I brought Vitaly to the train and picked up Scott, The Displex cooled down to 150 K and the beamline warmed up (hopefully). Aligning the sample and taking the first EXAFS scan:
 

Filename sample comments regions stepsize
19Nov00.014 V2O3
608 E8 Rest
motor scan Y
015 motor scan Z
016 EXAFS scan
E par. c (?)
150 K
017 dito
retuning optics, realigning sample
018 dito
Glitch at 5679.6 and 5901

going to 133 K, retuning optics (didn't change) and realigning sample
 

Filename sample comments regions stepsize
19Nov00.019 V2O3
608 E8 Rest
motor scan Y
020 motor scan Z
021 EXAFS scan
E par. c (?)
133 K
022 dito

going to 130 K, retuning optics (didn't change) and realigning sample
 

Filename sample comments regions stepsize
19Nov00.023 V2O3
608 E8 Rest
motor scan Y
024 motor scan Z
025 EXAFS scan
E par. c (?)
130 K
026 dito

going to 127 K and realigning sample
 

Filename sample comments regions stepsize
19Nov00.027 V2O3
608 E8 Rest
motor scan Y
028 motor scan Z
029 EXAFS scan
E par. c (?)
127 K
injection @19.00
030 motor scan Y
031 motor scan Z
032 EXAFS scan
E par c (?)
127 K
ugly Io due to moving source after injection. Probably scan is unusable.
aligning sample
033 dito, looks noisy, too
aligned sample
034 dito

calibration: two subpeaks in calibration region; chose higher one. Recalibrated 5454.5 to 5475.
Aligned mirrors.
Set to 124K (dropped to 123.7 while setting)
 

Filename sample comments regions stepsize
19Nov00.035 EXAFS scan
E par c (?)
124 K
Recalibrated 5447 to 5775
Realigned mirrors, but they seem to be stabilizing
036 dito, unusably noisy
realigned sample
037 dito, looky much better at the first glance, but not at second
038 next try, worse again. Killed somewhere in the middle
aligning sample:
039 motor scan Y
040 motor scan Z
041 EXAFS scan, again
looks much better, but different than all the previous ones. So I hav to take another one:
042 EXAFS scan, 124 K

 

2000-11-20

Still there on a new day!
Finally got two scans for 124 K that look alike and finishing that temperature that really gave me a hard time!

Going to 120 K
 

Filename sample comments regions stepsize
20Nov00.001 V2O3
608 E8 Rest
motor scan Y
002 motor scan Z
003 EXAFS scan
E par. c (?)
120 K
realigning sample
004 dito
Glitch at 6000.5eV

Strange things happen: Both scans are usable!
Changing temp to 116 K, doing the usual tuning and alignment:
 

Filename sample comments regions stepsize
20Nov00.005 V2O3
608 E8 Rest
motor scan Y
006 motor scan Z
007 EXAFS scan
E par. c (?)
116 K
sample realigned
008 dito
Glitch at 5676.8?
retuning optics, realigning sample
009 dito
010 2nd scan in set without recalibration

Next morning:
Went to the beamline befor breakfast to warm up the mirrors and change the temperature of the sample to 112 K
 

Filename sample comments regions stepsize
20Nov00.011 V2O3
608 E8 Rest
motor scan Y -2, 2 .1
012 motor scan Z
013 EXAFS scan
E par. c (?)
112 K
realigned sample
014 dito
015 motor scan Y
016 motor scan Z
017 EXAFS scan
E par. c (?)
112 K
looks glitchy, but isn't
018 dito
glitches at 5623 and 5682?!

Once it's working, it's really great: got two identical scans up to 14 A^-1 !
Going to 90 K now:
 

Filename sample comments regions stepsize
20Nov00.019 V2O3
608 E8 Rest
motor scan Y
020 motor scan Z
021 EXAFS scan
E par c (?)
90 K
looks good, but large E shift
checked alignment of sample. Did not move
022 dito
023 dito

Going down to 30 K
 

Filename sample comments regions stepsize
20Nov00.024 V2O3
608 E8 Rest
motor scan Y
025 motor scan Z
026 EXAFS scan
E par. c (?)
30 K
Glitch at 5991.5
checking alignment (didn't change)
027 dito
realigning sample
028 dito
Glitch at 5676.6

Finally heating to 150 K again:
 

Filename sample comments regions stepsize
20Nov00.029 V2O3
608 E8 Rest
motor scan Y
030 motor scan Z
031 EXAFS scan
E par. c (?)
150 K

FINISHED WITH V2O3 FOR THIS RUN!!!

Checking energy calibration of monochromator now, using Ti, V and Cr foil:
 

Filename sample comments regions stepsize
20Nov00.032 Ti foil
V foil
Cr foil
edge scans -700, -550, -450, -50, 50, 475, 575 5, .5, 5, .5, 5, .5
033 V foil
no Cr foil
Mn fiol
edge scan V
removed Cr instead of Ti foil, $%&*+!§
-100, 100 2
034 scanning all three edges
well, no Cr edge appeared ...
-50, 50, 475, 575, 1025, 1125 .5, 5, .5, 5, .5
035 V, Cr, Mn foils
(really this time!)
V edge scan as 033
036 all three edges now as 034

 
edge recorded position expected position drift
V 5462 5465 -3
Cr 5970.5 5989 -18.5 / 524eV
Mn 6499.5 6539 -39.5 / 550+524eV

 
Filename sample comments regions stepsize
20Nov00.037 V, Cr, Mn foil V edge scan
edge was at 5445 (?); calibrated it to 5465
-100, 100 2
038 all three edges -50, 50, 475, 575, 1025, 1125 1, 10, 1, 10, 1

 
edge recorded position expected position drift
V 5463 5465 -2
Cr 5969 5989 -20 / 524eV
Mn 6500 6539 -39 / 550+524eV

 
Filename sample comments regions stepsize
20Nov00.039 V, Cr, Mn foil V edge scan
edge was at 5449; calibrated it to 5465
-100, 100 2
040 EXAFS scan similar to the ones for V2O3 -200, -30, 30, 17k 2, .5, .05k

 
edge recorded position expected position drift
V 5459.5 5465 -5.5
Cr 5965 5989 -24 / 524eV
Mn 6500 6539 -39 / 550+524eV

 

Putting in the Mn foil.
 

Filename sample comments regions stepsize
20Nov00.041 Mn foil motor scan Z at 300 eV above Mn edge -4, 4 .1
042 Y .2
decreased gain on I_t to 1e8
043 edge scan -100, 100 2
044 Mn foil
Co reference
EXAFS scan -200, -30, 30, 18k 2, .5, .05k
045 garbage
046 checking reference detector 1120, 1220 5
047 2
048 turned I_ref "chamber" around
049 Mn foil
reference detector w/out foil
050 Mn foil and Co foil between Io and I_t detectors reference signal without foil
051 1020, 1220
found the Co edge 170 eV off and with an edge jump of 4!
052 quick scan through whole region -200, 17.5k 5
edge was too high, this time --> recalibrated
053 EXAFS scan -200, -30, 30, 18k 2, .5, .05k
054 Co foil only Scan as planned for Mn foil to check if Co signal is really smooth in th desired energy range
using integration time 1 for this scan to speed things up
LOST LOCK!!!
-200, -50, 50, 16.5k, 18k 5, 1, .05k, 2
055 dito
LOST LOCK!!!
056 Co edge scan 1020, 1220 5
057 like 054 5,5,5,5
058 garbage

 
 

2000-11-21
 

21Nov00.001 Mn foil and Co foil starting real (!) EXAFS scan on Mn now
LOST LOCK!!!
-200, -50, 50, 16.5k, 18k 5, 1, .05k, 2
retuned mirror 2 at 450eV above edge
increased sensitivity on lockin to 500mV from 1V
hopefully that helps and doe not make things worse!
002 another EXAFS scan
LOST LOCK!!!
dito dito
Adjusted mirrors at 800 eV above edge.
"Lowered" sensitivity to 100 mV.
Checked by moving monochromator
directly up and down...seems OK!
003 another EXAFS scan
worked--yay!
-200,-50,50,
16.5k,18k
5,1,.05k,2
004- more EXAFS scans
edge far off!
-200,-50,50,
16.5k,18k
5,1,.05k,2

returned next morning:
 

Filename sample comments regions stepsize
21Nov00.006 Mn foil + Co foil edge scan -200, 0 ?
recalibrated Mn edge from 6434 eV
007 EXAFS scan
nearly lost lock, but recovered after one point
-200, -50, 50, 16.5k, 18k 5, 1, .05k, 2
reaclibrated Mn edge (was approx. 20eV low)
trying a even higher sensitivity of 50mV (which corresponds to range (V/div) on oszi)
optimizing the flux 800 eV above the edge leads to a very low Io at the begin of the scan (20% of the value at +800eV). I'm trying +600eV. Does not change anything.
set bias to 0 at -200eV
008 new EXAFS scan at 250K
have been moving back and forth ca couple of times: edge is ~40eV too low.
LOST LOCK!!!
returning to sensitivity of 100mV
009 EXAFS scan, again.
LOST LOCK
as "rocking" sine curve was bumping around at high energies, I changed the gain a bit. the old position is marked with electrical tape (turn counterclockwise to return). Trying another scan
010 EXAFS scan
LOST LOCK AGAIN.
recalibrated edge (was at 6484)
011 quick scan with integration times of 1 sec to check it again. Did not set bias to 0 at begin of scan this time.
Went through.
changed gain a bit, retuned optics at +700eV and set bias to 0 there
012 doing an EXAFS scan with 2 sec integration time.
Worked! 
Probably setting the bias at 700eV above the edge (and not at the beginning of the scan) is the clue???

Reducing temp to 200 K.
 

Filename sample comments regions stepsize
21Nov00.013 Mn foil + Co foil EXAFS scan
200 K
014 dito

Reducing temp to 150 K.
 

Filename sample comments regions stepsize
21Nov00.015 Mn foil + Co foil EXAFS scan
150 K
016 dito

Reducing temp to 100 K.
 

Filename sample comments regions stepsize
21Nov00.017 Mn foil + Co foil EXAFS scan
100 K
018 garbage?
019 100 K EXAFS scan

 

Reducing temp to 50 K.
 

Filename sample comments regions stepsize
21Nov00.020 Mn foil + Co foil EXAFS scan
50 K
50 K EXAFS scan

Returned from dinner, put in Cr foil and Fe as a reference (16.7A^-1 above Cr edge).
Tried to tune optics. Flux was very low, so I tried to tune it for the V K edge where I knew the flux had to be around 1.0; was .01 what did not add to my pleasure. Noticed we had run out of He. Probably that already caused the flux to be so low during the Mn scans. Got a new bottle of compressed He for 34.91$ from the Stockroom.
Flux is slowly increasing.
Changing gain to 500mV

Starting 1st scan while gas flow still is settling
 

Filename sample comments regions stepsize
21Nov00.022 Cr foil + Fe foil EXAFS scan
300 K
1 sec integration times
-200, -50, 50, 16.0k, 17.5 k 5, 1, .05k, 2
023 EXAFS scan
300 K
2 sec integration
024 dito
LOST LOCK!
changing sensitivity to 1V (lockin is in overload)
025 new EXAFS scan
300 K

Lowering temperature to 250 K.
 

Filename sample comments regions stepsize
22Nov00.001 Cr foil + Fe foil EXAFS scan -200,-50,50,16.0k,17.5k 5,1,.05k,2
002 dito
Glitch in Io at 6967
adjusted bias at "edge"
(actually above because
of drift)
Moved 50 below the edge
and recalibrated as the edge
003 Cr foil + Fe foil 200 K EXAFS scan -200,-50,50,16.0k,17.5k 5,1,.05k,2
004 ditto
Before scanning, moved 20 eV
below edge and recalibrated.
005 Cr foil + Fe foil 150 K EXAFS scan -200,-50,50,16.0k,17.5k 5,1,.05k,2
006 trash
007 trash
recording last scan at 150 K the next morning:
008 150 K EXAFS scan
I_t detector may be saturated or nonlinear responding (went to ~11V, full saturation occurs at ~12V) above 6900eV (~15A^-1)

Starting 100 K scans:
 

Filename sample comments regions stepsize
22Nov00.009+010 Cr foil + Fe foil EXAFS scans at 100 K
#010 lost lock at 6964
011 dito

Starting 50 K scans:
 

Filename sample comments regions stepsize
22Nov00.012 Cr foil + Fe foil EXAFS scans at 050 K
wrong user comments in data file (still says 100 K)
013 dito

 
Filename sample comments regions stepsize
22Nov00.014 _NO_ Cr foil + Fe foil EXAFS scan on Cr edge with Fe foil only to chekc smoothness of Fe foil in Cr region
integration times set to 1 sec
-200, -50,50, 16k, 17.5k 5,2,.05k,2